Description
The Milwaukee MC120 PRO pH Monitor (MC120US) gives a continuous pH reading on a large LCD and flashes an alarm when pH goes above or below the set point you pick, anywhere from 5.5 to 9.5 pH. It comes with the MA911B/2 gel-filled, double junction pH probe on a 6.6 ft (2 m) cable, a 12 VDC adapter and a mounting kit with probe holders. Milwaukee lists it under its aquaculture and coral applications.
Specs
- Model: MC120 PRO pH Monitor, part number MC120US
- pH range: 0.0 to 14.0 pH
- Resolution: 0.1 pH
- Accuracy: ±0.2 pH
- Calibration: manual, 2 points, with trimmers on the front panel
- Alarm set point: 5.5 to 9.5 pH, alarm when the reading is higher or lower than the set point
- Probe: MA911B/2 double junction, gel-filled, polypropylene resin body, 6.6 ft (2 m) cable (included)
- Power: 12 VDC adapter (included)
- Dimensions: 5.9 x 3.2 x 1.35 in
- Operating environment: 32 to 122 °F (0 to 50 °C), max 95% RH
- In the box: monitor, MA911B/2 pH probe, 12 VDC adapter, mounting kit with probe holders
How to use
- Connect the probe to the BNC socket on the bottom of the meter, take off the protective cap and put the bottom 1 in (2.5 cm) of the probe in the water.
- Calibrate in pH 7 solution first: let the reading settle and turn the pH7 trimmer until the display reads 7.0. Then put the probe in pH 4 solution and turn the pH4/10 trimmer to 4.0.
- Set the alarm with the center front knob, then set the switch to ABOVE or BELOW. The alarm blinks when pH crosses your set point.
- Milwaukee suggests recalibrating at least once a month.
- Store the probe in MA9015 storage solution, never in distilled, DI or RO water. Soak it in MA9016 cleaning solution for about 20 minutes on a regular schedule.
Works well with
- Milwaukee MA911B/2 Double Junction pH Probe
- Hanna pH 7.01 Calibration Buffer Sachets (HI70007P)
- Milwaukee pH 4.01 Calibration Buffer 230 mL
- Milwaukee pH/ORP Probe Storage Solution 230 mL
Common questions
Does it come with a pH probe? Yes. The MA911B/2 double junction, gel-filled probe with a 6.6 ft (2 m) cable is included, along with the 12 VDC adapter and a mounting kit with probe holders.
Which calibration solutions do I need? Milwaukee’s manual calibrates in pH 7 solution first, then pH 4 solution, using the two trimmers on the front panel. Milwaukee suggests recalibrating at least once a month.
How should I store the probe? Keep it hydrated in MA9015 storage solution. Milwaukee warns never to store it in distilled, DI or RO water because that damages the probe’s sensitivity.
Questions? Ask our Reef Bot or call the shop at (305) 850-9949.
See more in Testing.
Shipping & Returns
- Equipment and dry goods ship at their own rate. Free shipping at $299 or more applies to qualifying livestock only — corals, fish, inverts and live copepods. The $5 Box & Handling Fee is a livestock packing fee and does not apply to dry-goods orders.
- All sales are final on equipment and dry goods. We are a retail store, not the manufacturer.
- Register your product with the manufacturer and keep your receipt — their warranty is how a faulty unit gets replaced, and they are faster at it than a shop could be.
- Not sure who to contact for a warranty claim? Call us and we will point you to the right place.
Questions? Call (305) 850-9949